Description
The ProSolo ODO/CT kit includes the digital ProSolo handheld; an optical dissolved oxygen, conductivity, and temperature cable/probe assembly (4 or 10 meter cable length); and a soft-sided carrying case.The ProSolo handheld is the direct replacement for the ProODO Optical Dissolved Oxygen handheld.
-
Overview
ODO® Sensor Technology Dissolved oxygen measurement technology has advanced from electrochemical sensors, such as polarographic and galvanic, to luminescent-based, optical sensors. YSI optical DO sensors are non-consumptive, meaning oxygen is not consumed during the measurement. As a result, there is no flow dependence or stirring required with ODO, allowing for the most accurate readings in the field.
Advantages over electrochemical sensors include:
Higher accuracy with no stir dependence
Less drift over time; holds calibration longer
Less maintenance; no membranes or electrolyte Real-Time Salinity Compensation
With an integral conductivity module, the ODO/CT probe assembly enables the ProSolo to automatically compensate DO mg/L measurements for changes in salinity. The 4-electrode conductivity cell, designed for rugged field performance, is easily cleaned and requires minimal maintenance. -
Technical Info
ProDIGITAL Smart Sensors
In addition to utilizing the most advanced DO sensing technology, the ProSolo incorporates smart sensor technology that enables sensors to be automatically recognized by the handheld. Additionally, DIGITAL sensors store calibration data, so probe assemblies can be swapped between handhelds without the need for re-calibration.
Provides Years of Sampling for All Field Conditions
The ProSolo handheld meets the demands of true field work with a rugged, waterproof case (IP-67 rated) and a military-spec (MS) cable connector. Rigorously drop tested from all angles, the handheld features a 3-year warranty, while a 2-year warranty is included with the probe assembly and ODO Cap.
Advanced Data Management
ProSolo includes KorDSS Software, a powerful data management program for easy 2-way communication between the meter and a PC. A unique data backup feature allows for the export of data to a USB flash drive (included) in csv format without having to connect to a computer.
Convenient Sampling Experience
Ergonomic handheld for a comfortable grip
Easy-to-read color display and backlit keypad
Large onboard memory with storage capacity for over 100,000 data sets
Site ID and Data ID tag capabilities for data organization
Data export directly to USB flash drive (included)
Single cable design for probe assembly lengths ranging from 1- to 100-meters
Long-life rechargeable lithium-ion battery that is user-replaceable
Simple and quick calibration procedures -
Size 2.46 cm diameter, cable options for 1, 4, 10, 20, 30, 50, and 100 m lengths Dissolved Oxygen
(Optical luminescence)
Units: % saturation, % saturation
local, mg/L, ppmRange: 0 to 500%, 0 to 50 mg/L
Accuracy: 0 to 200%: ±1% of reading or 1% saturation, whichever is greater
200 to 500%: ±8% of reading
0 to 20 mg/L: ±0.1 mg/L or 1% of reading, whichever is greater
20 to 50 mg/L: ±8% of reading
Resolution: 0.01 mg/L and 0.1%, or 0.1 mg/L and 1% (auto-adjusts based on range)
Temperature
(Thermistor)
Units: °C, °F, KRange: -5 to 70°C (temperature compensation range for DO mg/L measurement: -5 to 50°C )
Accuracy: ±0.2°C
Resolution: 0.1°C or 0.1°F (auto-adjusts based on range)
Conductivity
(Four nickel electrode cell)
Units: μS/cm, mS/cmRange: 0 to 200 mS/cm
Accuracy: 0 to 100 mS/cm ±0.5% of reading or .001 mS/cm, whichever is greater
100 to 200 mS/cm ±1.0% of reading
Resolution: 0.001, 0.01, or 0.1 mS (range dependent)
Specific Conductance
Units: μS/cm, mS/cm
Range: 0 to 200 mS/cm
Accuracy: 0 to 100 mS/cm ±0.5% of reading or .001 mS/cm, whichever is greater
100 to 200 mS/cm ±1.0% of reading
User selectable reference temperature (15 to 25°C; default 25°C) and compensation
coefficient (0 to 4%/°C; default 1.91%)Resolution: 0.001, 0.01, 0.1 mS/cm
Warranty 2 years on cable, probe, and ODO sensor cap
Specifications
- Optical DO: Optical DO
- Dissolved Oxygen: Dissolved Oxygen
- Temperature: Temperature